The built-in measuring system measures the activity in the syringe directly during the dose preparation process and displays it on the control PC screen.

The activity remaining in the mother vial is calculated and displayed similarly.

Completely eliminates the contact of hands with the syringe during the activity preparation.

The position of the syringe piston is controlled mechanically - manually.

Measuring Unit PT468R3

Technical Specifications

  • Range of doses ..... 25 MBq to 15 GBq for 99mTc
  • Accuracy ..... ± 3 % for 99mTc
  • The minimum volume of radiopharmaceutical should be 0,5ml in 5ml syringe and 0,3 ml in 3,0 ml syringe to achieve the optimal accuracy
  • After determining the calibration factors, it can be used for other gamma radiation radiopharmaceuticals with energy from 80 keV to 400 keV.

Accessories

  • PT643R0 Slanted stand for elution vial
  • PT500R2 Slanted stand for activity distribution
  • PT644R0 Syringe shield 3 ml (red)
  • PT645R0 Syringe shield 5 ml (blue)
  • PT646R0 Syringe shield 10 ml (green)
PT643R0 Slanted stand for elution vial

PT643R0 Slanted stand for elution vial

PT500R2 Slanted stand for activity distribution

PT500R2 Slanted stand for activity distribution

Dimensions

  • Height ..... 345 mm
  • Width ..... 235 mm
  • Length ..... 264 mm
  • Weight ..... 22 kg

Intuitive and user-friendly software

  • SW continuously displays the remaining activity in the vial (calculated value).
  • Linearity and stability tests - check emitter incorporated in the device accessories.
  • Reliable identification of different RF
  • SW recommends a value of individual dose according to the patient’s weight or BMI based on present specifications.
intuitive and user friendly software pt468

Consumables for One Vial

  • 1x BBraun Spinocan 0,90 x 88mm; G20
  • 1x BBraun Sterican 0,80 x 120mm; G21
automatic dispensing unit pt317

eu footer