The built-in measuring system measures the activity in the syringe directly during the dose preparation process and displays it on the control PC screen.
The activity remaining in mother vial is calculated and displayed similarly.
The position of the syringe piston is controlled by operators hand.
Easy daily QC procedure as one of the SW features.
Technical Specifications
- Range of doses ..... 25 MBq to 15 GBq for 99mTc
- Accuracy ..... ± 3 % for 99mTc
- Dose can be prepared into 1 ml, 3 ml, 5 ml and 10 ml syringe
- The minimum volume of radiopharmaceutical should be 0,5 ml in 5 ml syringe and 0,3 ml in 3 ml syringe to achieve the optimal accuracy
- After determining the calibration factors, it can be used for other gamma radiation radiopharmaceuticals with energy from 80 keV to 400 keV.
Accessories
- PT490R2 Slanted stand for elution vial
- PT194R4 Slanted stand for activity distribution
- PT495R1 Syringe shield 3 ml
- PT494R2 Syringe shield 5 ml
- PT646R0 Syringe shield 10 ml
PT490R2 Slanted stand for elution vial
PT194R4 Slanted stand for activity distribution
Dimensions
- Height ..... 160 mm
- Width ..... 103 mm
- Length ..... 234 mm
- Weight ..... 3,2 kg
Intuitive and user-friendly software
- Continuously displays the remaining activity in the vial (calculated value).
- Linearity and stability tests - check emitter incorporated in the device accessories.
- Reliable identification of different RF
- The software will recommend a value of individual dose according to the patient’s weight or BMI based on preset specifications
Consumables for One Vial
- Only standard needles